User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Sub Fluke 83,85,87 Capacitance (1 year) CAL VER DATE: 26-Sep-97 AUTHOR: User Contributed REVISION: 2.1 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 49 STANDARD: GenRad 1409-Y STANDARD: GenRad 1409-X STANDARD: GenRad 1409-R STANDARD: GenRad 1409-K ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P J F W 1.002 ASK+ K X 1.003 DISP Press the BLUE button to select Capacitance. 1.004 DISP Connect 1µF between the V Ohm Diode and COM inputs. 1.005 STD GenRad 1409-Y 1.006 ACC 1.00µF 0.1% 1.007 MEMI Use keyboard to enter actual UUT reading in µF: 1.008 MEME 1.009 MEMC µF 1% 0.02U 5.0µF_Range #! Test Tol 0.03, Sys Tol 0.0075, TUR 4.000 (>= 4.00). 2.001 DISP Connect 0.5µF between the V Ohm Diode and COM inputs. 2.002 STD GenRad 1409-X 2.003 ACC 0.500µF 0.1% 2.004 MEMI Use keyboard to enter actual UUT reading in µF: 2.005 MEME 2.006 MEMC µF 1% 0.002U 0.5µF_Range #! Test Tol 0.0067, Sys Tol 0.001645, TUR 4.073 (>= 4.00). 3.001 DISP Connect 0.05µF between the V Ohm Diode and COM inputs. 3.002 STD GenRad 1409-R 3.003 ACC 0.0500µF 0.1% 3.004 MEMI Use keyboard to enter actual UUT reading in µF: 3.005 MEME 3.006 MEMC µF 1% 0.0002U 0.05µF_Range #! Test Tol 0.00067, Sys Tol 0.0001645, TUR 4.073 (>= 4.00). 4.001 DISP Connect .005µF between the V Ohm Diode and COM inputs. 4.002 STD GenRad 1409-K 4.003 ACC 5.00nF 0.1% 4.004 MEMI Use keyboard to enter actual UUT reading in nF: 4.005 MEME 4.006 MEMC nF 1% 0.35U 5.0nF_Range #! Test Tol 0.397, Sys Tol 0.0235, TUR 16.894 (>= 4.00). 5.001 DISP Remove all connections. 5.002 HEAD 5.003 END